Greece
Menu

Zenit fast

The information to be found in this section is intended for the exclusive use of healthcare professionals.

Magnification
4x, 20x
Optional magnification
Three extra positions to introduce standard microscope objectives, up to five loaded objectives. Maximum magnification 40x
Scanning speed
Less than 30 seconds/well for ANA HEp-2 slides at 20x
Less than 2.5 minutes/well for tissue slides at 20x
Resolution at 20x
0.5 micron
Scanning mode
Single or multiple runs per session
Type of compression
Standard JPEG-2000

Compression level
ANA/ANCA/nDNA: 5-10 MB per well; tissues: 100-150 MB per well

2D Data matrix reader
Yes

Connectivity
Bidirectional, HL7 protocol

Scanning capability
HEp-2, ANCA, nDNA, tissues

Automated analysis
Zenit ANA HEp-2 pattern recognition (8 patterns identified): AC-0 Negative, AC-1 Nuclear homogeneous, AC-2,4,5,29 Speckled, AC-3 Centromere, AC-6,7 Nuclear dots, AC-8,9,10 Nucleolar, AC-15,16,17,18,19,20,22,23 Cytoplasmic, AC-21 Cytoplasmic reticular AMA
Dedicated gallery of mitoses on recognized patterns and relocalization

Zenit ANCA ethanol and ANCA formalin pattern recognition (Negative, c-ANCA, p-ANCA, Others)
Zenit nDNA positive/negative recognition

Atlas / Pattern guide
Reference images with ANA Nomenclature ICAP codes

Electrical specifications
Power source: AC 100-240V 50/60 Hz
Power consumption: 45 VA

Mechanical specifications
Scanner: Height: 50 cm / Width: 60 cm / Length: 40 cm / Weight: 30 kg
Processing unit: Height: 46 cm / Width: 52 cm / Length: 21 cm / Weight: 15 kg
Display unit: Inches: 24” / Resolution: 1920x1200 / Height: 46-55 cm / Width: 25 cm / Length: 57 cm / Weight: 10 kg